Measurement accuracy: internal check
A distance measured in the viewer is only as good as the way the software reads pixel size, slice position and orientation from the DICOM file. We check this regularly with synthetic test studies of known dimensions. This page explains what we measure against, how, and what we found.
Important: Sinus3D is not a medical device. Display and measurements are for information only and not for diagnostic use. This is an internal test, not an external certification. For diagnosis, treatment planning and measurement, download the original DICOM and open it in software qualified for that purpose.
What do we measure against?
On a real patient study the “correct” distance is unknown, so we build synthetic test studies (phantoms) in which every dimension is known exactly. Each contains five small dense markers at known positions and a block of known density (300 HU) in an air-like (−1000 HU) background. No patient data is used.
The phantoms cover 11 study geometries seen in practice: dental CBCT, hospital CT with thick and overlapping slices, MR series in different planes, oblique orientation, gantry tilt, a missing slice and multi-frame files.
How do we measure?
- Two paths, just as you use it: (a) uploaded — the server prepares the study for the viewer; (b) in the local viewer, in the browser, without upload.
- In the viewer we locate the centre of each marker (from its full image, with sub-pixel precision) and compute the ten distances between the markers using the viewer’s own pixel sizes — the same formula as the Measure tool.
- We compare these with the true, known distances.
- We also read the density of the block (HU).
The test runs automatically and is repeated after every change to the viewer or the converter.
Results
For all 11 study types, both paths gave the same result:
| Study type | Voxel size (mm) | Largest deviation |
|---|---|---|
| Dental CBCT, isotropic | 0.2 × 0.2 × 0.2 | 0.000 mm |
| Hospital CT, thick slices | 0.7 × 0.7 × 3 | 0.030 mm |
| CT overlapping reconstruction (2 mm thick, 1 mm spacing) | 0.7 × 0.7 × 1 | 0.000 mm |
| Non-square pixels | 0.5 × 0.8 × 1 | 0.019 mm |
| Reversed slice order | 0.6 × 0.6 × 1.25 | 0.012 mm |
| Sagittal MR series | 0.9 × 0.9 × 4 | 0.011 mm |
| Coronal series | 0.8 × 0.8 × 2 | 0.043 mm |
| Oblique orientation (20°) | 0.6 × 0.6 × 1 | 0.034 mm |
| Gantry tilt (15°) * | 0.7 × 0.7 × 1.45 | 0.015 mm |
| Missing slice * | 0.7 × 0.7 × 1.5 | 0.012 mm |
| Multi-frame (enhanced) DICOM | 0.3 × 0.3 × 0.3 | 0.006 mm |
The largest deviation is 0.043 mm — a fraction of the pixel size, due to numerical rounding. The 300 HU block was read as exactly 300 HU in every case.
* During the test we found and fixed two issues (24 November 2026):
- Gantry tilt (mainly older head CTs): tilting shifts each slice within its own plane. The viewer did not account for this, which caused up to 6 mm deviation across slices and a distorted side view. The slices are now realigned when the volume is assembled.
- Missing or uneven slice spacing: the viewer used the average spacing (0.5 mm deviation in the test). Slices are now resampled to an even grid.
What this test does not cover
- Click precision: you place the measurement endpoints on the image — roughly half a pixel of uncertainty per endpoint (about ±0.1 mm on CBCT; up to ±1.5 mm across slices on thick-slice CT). Zooming in reduces it.
- Physical limits of the study: thick slices blur the edges of small structures; noise and motion affect the image itself — no viewer can correct these.
- Scanner data: if the scanner writes a wrong pixel size into the DICOM file, every viewer — ours included — measures accordingly.
- Distances along the curved plane of the panoramic reconstruction are not yet part of this test; they are planned for the next check.
Questions?
If you notice a discrepancy on a study, or would like a more detailed description of the method, write to info@sinus3d.com. See also: How to view studies · About us